3.6.2.1
User defined events
The monitoring functions of the device can be configured for user defined events. By default, events are deactivated
(action = NONE). Contrary to device alarms, the events only work while the DC input is switched on. It means, for
instance, that you cannot detect undervoltage (UVD) anymore after switching the DC input off and the voltage is
still sinking.
The following events can be configured independently and can, in each case, trigger the actions NONE, SIGNAL,
WARNING or ALARM.
Event
Meaning
UVD
UnderVoltage Detection
OVD
OverVoltage Detection
UCD
UnderCurrent Detection
OCD
OverCurrent Detection
OPD
OverPower Detection
These events should not be confused with alarms such as OT and OVP which are for device
protection. User defined events can, however, if set to action ALARM, switch off the DC input
and thus protect the source (power supply, battery)
► How to configure user defined events
1.
Tap the touch area
2.
On the right side tap the arrows
3.
Set the monitoring limits with the left hand rotary knob and the triggered action with the right hand knob
relevant to the application (also see „3.6.1. Definition of terms").
4.
Accept the settings with
As soon as an event is set up with an action other than "NONE" and with accepted settings, an incident can occur
whether the DC input is switched on or off. On leaving the pages "User events" or "Settings" an event can be
directly displayed.
User events are an integral part of the actual user profile. Thus, if another user profile, or the
default profile, is selected and used, the events will be either differently or not configured.
The set values can be entered using the ten-key tab. This will appear by tapping the touch area
for direct input (ten key pad symbol) on the particular page, e.g. "4.1 Event U".
EA Elektro-Automatik GmbH
Helmholtzstr. 31-33 • 41747 Viersen
Germany
Description
Triggers an event if the input voltage falls below the
defined threshold.
Triggers an event if the input voltage exceeds the de-
fined threshold.
Triggers an event if the input current falls below the
defined threshold.
Triggers an event if the input current exceeds the de-
fined threshold.
Triggers an event if the input power exceeds the defined
threshold.
on the main screen.
to select "4.1 Event U" or "4.2 Event I" or "4.3 Event P".
.
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
ELR 9000 Series
www.elektroautomatik.de
ea1974@elektroautomatik.de
Range
0 V...U
Nom
0 V...U
Nom
0 A...I
Nom
0 A...I
Nom
0 W...P
Nom
Page 65