outPut waveFoRm:
Quasi Sinusoidal
electRical iSolation:
4kV
@ 60Hz from patient-connected parts to ground.
rms
imPedance:
Range: 15Ω to 999Ω
Rated Load: 50Ω
Accuracy: ±10% ±5Ω
(displays "LO" below 15Ω and "HI" above 999Ω)
time diSPlay:
Range: 0 to 20 minutes
Accuracy: ±1%
Resolution: 1s
RetuRn Pad cuRRent:
Current Limit to any pad: 0.79A
(P-X error, where X is the pad number)
mainS inPut voltaGe:
100-120V~ or 220-240V~ configured through the power entry module fuse block
and a line voltage switch.
mainS inPut FRequency:
50/60 Hz
inPut cuRRent:
4.8 A
max @ 100-120V~
rms
2.5 A
max @ 220-240V~
rms
FuSeS:
100-120V~: T6.3AL 250V, 5 x 20 mm
220-240V~: T3.15AL 250V, 5 x 20 mm
theRmal ShutoFF:
Excessive temperature (80 °C) inside the chassis inhibits RF output.
RS-232 PoRt:
Female DB9 connector, with the following pinout:
Pin2:
RxD
Pin3:
TxD
Pin5:
GND (serial communication reference)
Pins 1, 4, 6-9: n/c|
RS-232 Communication protocol: 9600 Baud, No Parity, 8 bits, 1 stop bit (9600, N, 8, 1)
Size:
35.5 cm W x 41.9 cm L x 10.6 cm H (14" W x 16.5" L x 4.2" H) (including protrusions)
weiGht:
6.350 kg (14 lb)
oPeRatinG conditionS:
Temperature: +10 °C to +40 °C
Humidity: 30% to 75% (non-condensing)
Atmospheric Pressure: ≥ 800 millibars (up to 2,000 meters)
CAUTION! For altitudes above 1,500 meters, the maximum operating
temperature is 35 °C.
StoRaGe/tRanSPoRt conditionS:
Temperature: 0 °C to +70 °C
Humidity: 0% to 95% (non-condensing)
Atmospheric Pressure: ≥ 800 millibars (up to 2,000 meters)
±5% ±0.02A
rms
rms
5
max.