Stop
RecePtacleS
Symbol
ReaR Panel
item
Stop – Transfers the system from Operating mode into
Standby mode.
Electrode Receptacle – Accepts the plug on the electrode
cord for energy delivery through the LeVeen™ electrode
or Soloist™ Electrodes.
Return Receptacles – Accept the plugs on the Megadyne
0850C Patient Return Electrodes. Also, uniquely identifies
the four return pads for troubleshooting.
Safety Symbol – Indicates that the receptacles have
CF (Cardiac, Floating) type electrical isolation from
mains and ground and that the RF 3000™ Generator is
protected from damage during a defibrillator discharge
directly into the receptacles.
Electrode isolated from earth ground at high frequency.
Earth Ground Stud – Provides a means of securely
linking the earth grounds of the RF generator to other
grounded equipment.
RS-232 Port – For authorized service personnel only.
Power Entry Module – This module contains both
the power ON/OFF switch and the fuses. The voltage
is selected by the orientation of the fuse drawer, as
marked.
Voltage Selector – This switch configures the generator
to either 110V~ or 220V~ in conjunction with the Power
Entry Module. The appropriate tool must be used to
change configurations.
Volume Control Knob – This knob controls the tone
volume from 40dBA @ 1 meter except pad-off tones
which are fixed at 65dBA @ 1 meter.
8