G
uiDance anD manufacturer
The RF 3000™ Radio Frequency Generator is intended for use in the electromagnetic
environment specified below. The customer or the user of the RF 3000 Radio Frequency
Generator should assure that it is used in such an environment.
i
mmunity test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
note - The RF 3000 does not possess a DC power port.
The RF 3000 interface cables do not exceed 3 meters in length.
'
s Declaration
iec 60601
c
ompliance level
test level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input/
±1 kV for input/
output lines
output lines
±1 kV
±1 kV
differential mode
differential mode
±2 kV
±2 kV
common mode
common mode
<5% Ut
<5% Ut
(>95% dip in Ut)
(>95% dip in Ut)
for 0.5 cycles
for 0.5 cycles
40% Ut
40% Ut
(60% dip in Ut)
(60% dip in Ut)
for 5 cycles
for 5 cycles
70% Ut
70% Ut
(30% dip in Ut)
(30% dip in Ut)
for 25 cycles
for 25 cycles
<5% Ut
<5% Ut
(>95% dip in Ut)
(>95% dip in Ut)
for 5 sec
for 5 sec
3 A/m
3 A/m
16
–
electromaGnetic immunity
e
lectromaGnetic
environment
GuiDance
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
RF 3000 Radio
Frequency Generator
requires continued
operation during
power mains
interruptions, it is
recommended that
the RF 3000 Radio
Frequency Generator
be powered from an
uninterruptible power
supply or battery.
Power frequency
magnetic fields
should be at levels
characteristic of
typical location in a
typical commercial or
hospital environment.
-