Herunterladen Inhalt Inhalt Diese Seite drucken

Boston Scientific RF 3000 Gebrauchsanweisung Seite 18

Hochfrequenz-generator
Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 39
G
uiDance anD manufacturer
The RF 3000™ Radio Frequency Generator is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user
of the RF 3000 Radio Frequency Generator can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the RF 3000 Radio Frequency Generator as recommended
below, according to the maximum output power of the communications equipment.
r
ateD
maximum
output power
150 khz to 80 mhz
of
transmitter
w
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
note 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
'
s Declaration
s
eparation Distance accorDinG to frequency of transmitter
80 mhz to 800 mhz
0.18
0.37
1.17
3.70
11.7
18
electromaGnetic immunity
m
800 mhz to 2.5 Ghz
0.18
0.37
1.17
3.70
11.7
0.23
0.74
2.33
7.37
23.3