laBelinG
Symbol
initial PoweR-on mode
CAUTION! Ensure that the Voltage Selector and the fuse block on the rear panel are
set to the appropriate mains voltage before turning the mains power switch on.
The RF 3000™ Generator is switched on by setting the mains power switch, located
on the rear of the unit on the power entry module, to the "|" position.
After mains power is applied, the RF 3000 Generator will briefly illuminate all
displays and sound a short tone. A self test is performed with a brief application
of RF power to an internal load. If any problems are identified, an error code will
be displayed in the power window and the unit goes to Fault mode. If the self test
is successful, the unit will proceed to the Ready mode.
Ready mode
In the Ready mode, the RF 3000 Generator is ready for ablation duration time
to be adjusted. The TIME window will display the default time of 15:00 minutes,
or a selected time in the acceptable range of values: 00:30 to 20:00 minutes.
RF power is not applied in this mode of operation, therefore the ACTIVE light and the
"RF On" tone are off. POWER and IMPEDANCE windows will be blank.
When START is pressed, the Operating mode is initiated.
Caution: Dangerous Voltage
Caution / Warning
Earth Ground
Fuses
Power ON
Power OFF
Radiofrequency energy source
Apply 2 return pads to each leg (four pads total)
Alternating Current
9
meaning