3.10.2 Settings
An initial setting is required before start of operation. This procedure has to be performed for
each sensor separately:
Connect sensor to display unit SEB2.2
Connect display unit SEB2.2 to PC
Start PinchPilot
In the menu go to "Sensor/ Configuration/ Extended"
Activate "Measurement output via Analog Interfaces"
Set maximum force: 250N, 500N, 1000N, 2000N
Push "Write" button
Repeat procedure for each sensor to be used.