3.5.3 Mode of measurement
Precondition: selectable mode (PinchPilot: sensor/configuration/extended).
Depending on the standard, a single, a dual or a triple measurement is required. An
additional setting is the option "auto sequence". In this case, the next measurement is started
automatically. The user can fully concentrate on the closing system to be tested.
Settings:
(requires activation via PinchPilot: Sensor/Configuration/extended: O selectable mode)
SEB2.2: After turning on the actual battery status is displayed for 2s. Press during this
phase the button again for 2s. Then the actual limit is displayed for 2s.
SEB2: After turning on the actual limit is displayed for 2s.
Press operating button again for two seconds. The actual mode is displayed for two
seconds.
Press briefly to switch to the next mode and repeat until the desired mode is set.
Display
1
Single measurement
Single measurement with auto sequence
+1
permanent restart, display of last peak force, no graphic data acquisition
2
Dual measurement
Dual measurement with auto sequence
+2
automatic start of second measurement
3
Triple measurement
Triple measurement with auto sequence
3+
automatic start of second and third measurement
5
Quintuple measurement
5+
Quintuple measurement with auto sequence
automatic start of second, third, fourth and fifth measurement
3.6
ERROR MESSAGES
If the display unit detects a malfunction, an error message is displayed (E:xx). The meaning
of the two digits error no. is explained in the appendix table "Error messages display unit".
3.7
DISPLAY UNIT MAINTENANCE
The display unit does not need any special maintenance. It should not be exposed to direct
sunlight for a longer time. Working range is temperature 0° to +40° Celsius at maximum 90%
humidity non-condensing. When dirty, clean it carefully with a soft, damp cloth. Avoid contact
to acids and thinner. For more information see section "Technical data".