The function of the operating button is different, according to the state of the display unit. In
the "off" state the display unit can be turned on. In the "on" state the display unit can be
turned off by pressing the button longer than two seconds.
By a short push on the operating button (with a connected sensor) a measurement is started.
Afterwards the result can be stored by a short push, or by a long push being rejected. When
not in measurement state the SEB2 can be switched off by a long push on the operating
button.
After initialisation, the display unit will enter the ready state. The state LED flashes slowly.
The number of the next free memory location is displayed.
In case of an expired calibration period (> 15 months), the red result LED flashes (refer to
chapter "General details/ Calibration").
Display of the next free memory no.
Type of measurement
Single measurement
Triple measurement
Triple measurement
Triple measurement
3.2
MEMORY CAPACITY
Measurement data is stored in a compressed format. The capacity of the memory is
app. 80 single or triple measurements.
3.3
DISPLAY UNIT EVALUATION
The peak force over the entire measurement period is displayed. This value is compared with
the permissible peak force. Additionally, the pulse duration is compared with the actual limit
value. Depending on this comparison, the red or green LED will be lit up.
Version 7.1
Handbuch / Manual FM200
Memory no.
Measurement no.
1
1
1
1
Seite 64 von 99
Display
1
.01.
1
.01.1
2
.01.2
3
.01.3