Classification
Protection against electric shock:
Class I Medical Electrical Equipment with Type BF defibrillation-proof applied parts
Protection against ingress of water: IPX1
Mode of operation: Continuous operation�
MEDICAL – GENERAL MEDICAL EQUIPMENT AS TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE
WITH ANSI/AAMI ES 60601-1 (2005) + AMD (2012), CAN/CSA-C22�2 No� 60601-1 (2008) + (2014), and IEC 60601-1-6:2010 (Third
Edition) + A1:2013; Control No� 4HZ8
Electrical characteristics
Leakage current
Meets leakage current requirements in accordance with IEC 60601 1�
Power cord
15 feet (4�6 m)
Device rating
110-120 VAC, 50/60 Hz, 1 Amp
220-240 VAC, 50/60 Hz, 0�8 Amp
Fuse
2 x F1A-H, rated 250V, for 110-120 VAC unit
2 x F0�8A-H, rated 250V, for 220-240 VAC unit
Storage and transport conditions
Storage/transport temperature
-20 to 60°C (-4°F to 140°F)
Store all components at room temperature and in a dry place when in use�
Operating humidity
Up to 90% RH, noncondensing
Atmospheric pressure range
Altitude up to 2000m or 80 kPa
Performance characteristics
Operating pressure
300 ± 10 mmHg setpoint
Note:
• Pressure system is In Range when the pressure infusor bladders are inflated to between 230 mmHg (low) and 330 mmHg (high)� If pressure falls below
230 mmHg for more than approximately 30 seconds the Low yellow indicator will illuminate and an audible indicator will sound� The High yellow and
audible indicator notifies the user when the pressure infusor bladder is above 330 mmHg�
• The outlet pressure of the fluid may vary with the surface area and volume of the fluid bag�
11