Introduction: Read the information contained in this brochure.
Understanding the information will assist you in using the MEDRAD
Avanta Bottle Heat Maintainer properly.
Important Safety Notice: The MEDRAD Avanta Bottle Heat
Maintainer is intended to be used by medical professionals with
adequate training and experience in angiographic procedures and
techniques.
Intended Use: The MEDRAD Avanta Bottle Heat Maintainer is
intended to maintain contrast heat in a contrast bottle for use in
diagnostic and interventional angiographic procedures performed in
cardiology, radiology, and vascular surgery, provided the operator has
pre-warmed the contrast.
Contraindications: Not intended for use with any other equipment
except the MEDRAD Avanta Fluid Management Injection System.
Restricted Sale: Rx Only.
Reach compliance information can be found at
www.REACH.bayer.com
Warnings
Burn Hazard: Operator injury may result if contact is
made with the heater element. Use caution when handling
the heat maintainer elements.
Patient or operator injury may result if package is opened
or damaged, or if damaged components are used. Visually
inspect contents and package before each use.
Cautions
Failure to use the correct connection port may damage
the heat maintainer connector. Properly identify the correct
connection port on the system before connecting.
MEDRAD Avanta Bottle Heat Maintainer
The MEDRAD Avanta Bottle Heat Maintainer should be used at the
discretion of the operator and is not required for proper functioning of
the MEDRAD Avanta Fluid Management Injection System.
The MEDRAD Avanta Bottle Heat Maintainer is available in two sizes,
a large size designed for 500 ml contrast bottles and a medium size
designed for 150 ml to 250 ml contrast bottles. These heat maintainers
are not guaranteed to fit all contrast bottles and configurations. Verify
that they fit securely on the contrast bottle before use.
Note: Stretching a medium bottle heat maintainer for use on a
500 ml bottle will result in improper heating.
Cleaning
The MEDRAD Avanta Bottle Heat Maintainer should be cleaned with
a dampened cloth using soap and water. Always disconnect power
before cleaning.
Fault Indicator
An illuminated red light indicates that the MEDRAD Avanta Bottle Heat
Maintainer has failed. Disconnect the heat maintainer from the system,
wait at least 30 seconds and then reconnect the heat maintainer. If the
red light remains illuminated, disconnect and discontinue use of the
MEDRAD Avanta Bottle Heat Maintainer. Call Bayer HealthCare
Services.
MEDRAD Avanta Bottle Heat Maintainer Installation:
®
50
AVA
0 BTLH-L
Grips
500 ml
1. Install the MEDRAD Avanta Bottle Heat Maintainer by sliding it up
from the bottom onto the contrast bottle.
Note: Ensure that the MEDRAD Avanta
Bottle Heat Maintainer is attached to the
contrast bottle in the correct orientation
as indicated by the arrows symbol.
Failure to do so may cause improper
heating of contrast.
2. Plug the MEDRAD Avanta Bottle Heat Maintainer into the
MEDRAD Avanta Bottle Heat Maintainer Port located on the back
of the MEDRAD Avanta Fluid Control Module.
Note: Do not use excessive force to engage the connector.
MEDRAD Avanta
Bottle Heat Maintainer Port
3. Check the MEDRAD Avanta Bottle Heat Maintainer for warmth
immediately after connection to the MEDRAD Avanta Fluid Control
Module.
6
AVA 500 BTLH-M
150 ml – 250 ml

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Medrad avanta avav 500 btlh-m

Inhaltsverzeichnis