6 Commissioning
88
Sample mode
Menu structure:
event program
Sample >
Distribution >
Parallel sample >
Esc
â
<-'
Fig. 37: Sample mode menu structure
The event programme contains the same sampling options (time, quantity, ext. signal
and flow) as those of the main and switching programmes. In addition, you can select
the function <ONCE>. In this function, the sampler takes a sample in the event
programme once and then returns immediately to the main programme if the event
signal is no longer present.
Distribution
You can reserve separate bottles for the event programme. With the exception of
parallel reserve sampling, the following applies when distributing to bottles:
The first bottle group of a distribution is reserved for the main programmes.
The second bottle group is reserved for the switching programmes.
The last bottle group is reserved for the event programme.
Menu structure:
event program
Sample >
Distribution >
Parallel sample >
Esc
â
á
Fig. 38: Distribution menu structure
You can define the bottles for the event programme as follows:
no change
move
1-9 bottles
Sample
-mode
:
time
time
:
volume
:
shots
:
Esc
â
á
Distribution
-mode
:
time
:
switch
: no change
<-'
Esc
â
There is no bottle change when the system switches to the event
programme.
The next empty bottle is filled when the system changes to the event
programme.
When the system switches to the event programme, 1-9 bottles are filled from
the last distribution bottle group. The number of bottles you can reserve for
the event programme is dependent on the total number of bottles (max.
9 bottles).
ASP station 2000
time
quantity
00:15 ext. sig.
100ml flow
01 once
<-'
time
time
quantity
01:00 ext. sig.
á
<-'
Endress+Hauser

Werbung

Inhaltsverzeichnis
loading