3 Installation

Installation conditions

64
3.3
Installation conditions
3.3.1

Dimensions

You can find the dimensions of the sampler in the Section "Mechanical construction" on
page 107.
3.3.2

Installation site

Fig. 3:
Notes on mounting the device
When mounting the device, please observe the following points:
❑ Position the device on a level surface.
❑ Protect the device against additional heat (e.g. radiators).
❑ Protect the device against mechanical vibration.
❑ Protect the device against strong magnetic fields.
❑ Ensure that air can circulate unhindered at the back of the cabinet. Do not position
the device directly against a wall (the distance between the wall and the rear side of
the cabinet should be at least 100 mm).
❑ Do not position the cabinet directly above the channel of the inflow to the wastewater
treatment plant (sulphurous vapours!).