ASP station 2000
Sampling
Inputs
Outputs
Memory interval
Thermostat
Interfaces
Creating programmes
Number
Creating programmes
Name
Creating programmes
Switch (only for an
active switching
Creating programmes
programme)
Endress+Hauser
Set
Time set.:
Basic settings
You can change the rinsing, suction, dosing and emptying
Sampling
times manually.
Delay:
You can delay the start of sampling (e.g. after an external
signal) by up to 99 seconds.
CS 1/2:
You can adapt the sensitivity of conductivity probes to different
media (e.g. high for media with poor conductivity).
Dosing:
Support dosing operation with pressure (e.g. for low suction
heights) or make unpressurised.
No flow:
Count and save if there is no inflow during sampling (e.g. with
blockages).
Set
Function for digital input 1, 2 and 3 and for setting the
Basic settings
analogue input.
Inputs
Set
Function for relay outputs 1, 2 and 3.
Basic settings
Outputs
Set
Time interval for internal recording of measured flow values. 0
Basic settings
sec = no recording.
Memory interval
Set
Configuration of thermostat in sample compartment; setting of
Basic settings
sample compartment temperature and defrost times.
Thermostat
Set
Transmission rate and definition of the RS232 interface
Basic settings
Interfaces
Set
1: 1 main programme active
2: 1 main programme and 1 switching programme active
Number
1+Event:
1 main programme and one event programme active
2+Event:
1 main programme, 1 switching programme and 1 event
programme active
Set
Enter names for the four main programmes, two switching
programmes and the event programme.
Name
Set
Day:
Switch between main and switching programmes on two
Switch
specific times a day.
Week:
Switch between main and switching programmes on two
specific weekdays.
Q greater:
Switch when flow greater than limit value.
Q smaller:
Switch when flow smaller than limit value.
External:
Switch on external signal.
6 Commissioning
Automatic
00 sec.
Standard
No press.
Count
No
Power down
0000 sec
04 °C
1
programme 1-4;
1-U1; 1-E;
1-U1-E
Day
91