5.6.2
Manual sample volume setting on standard device
Set the required sample volume by shifting the dosing pipe manually.
Calibrate the sample volume as follows:
Fig. 21: Sample volume calibration
1. Check the set sample volume of the active
programme.
2. Loosen the clamping lever and the air tube to remove
the dosing glass.
3. Pull the dosing glass forwards and remove it.
4. Unscrew the bayonet lock and open the dosing
glass.
5. Set the sample volume by moving the dosing pipe.
Then put the dosing glass back together and install it in
the reverse order.
5 Operation
75