Herunterladen Inhalt Inhalt Diese Seite drucken

Configuring The Internal Datalogger (Optional) - Endress+Hauser asp-station 2000 Betriebsanleitung

Vorschau ausblenden Andere Handbücher für asp-station 2000:
Inhaltsverzeichnis

Werbung

6 Commissioning
!
!
80
Analogue input
Menu structure:
Fig. 25: Analogue input menu structure
Signal
Units
Dec. point
Range
Note!
For the described functions, an analogue signal must be assigned at the analogue
input. For information on the correct wiring of the analog input see Section 4.1.5.
6.4.2

Configuring the internal datalogger (optional)

As an option, the ASP station 2000 can possess an internal datalogger. This saves the
data of a connected analogue signal and the sampling details (sample volume, bottle
filling times, events, etc.). The analogue value can be a quantity signal, but can also be
a pH value, conductivity, turbidity, etc.
Note!
When configuring the analogue input, note the SI unit of the analogue signal.
The minimum scan rate of the internal datalogger is 1 second. You can read off the
contents of the internal datalogger to a PC via the RS-232 interface using ReadWin2000
(see 5.5 "Communication using ReadWin® 2000").
Enter the output signal of the connected device: 0-1 V, 0-20 mA, 4-20 mA
Enter the SI unit of the analogue signal.
Enter the number of decimal places of the analogue signal.
Enter the maximum measuring range.
ASP station 2000
Endress+Hauser

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis