ASP station 2000
!

Configuring the inputs

!
Endress+Hauser
6.4

Device configuration

Note!
The figures show the menu structures with the display and picklist for the different
configuration parameters. Below the figures are tables containing the corresponding
function descriptions for the parameters.
6.4.1
Configuring the inputs
You can assign the inputs when configuring in the "Quick Setup" (see Section 6.3).

Digital inputs

Menu structure:

Fig. 24: Digital input menu structure
Sampling
Bottle
Quantity
Ext. stop
Event
Report
NC contact
NO contact
Switch
Note!
For the described functions, a digital signal must be assigned to the digital input 1, 2 or
3. For information on the correct wiring of the digital inputs see Section 4.1.5.
The input signal triggers a sample.
The input signal triggers a switch to the next empty bottle.
The input signal is a pulse signal from a volume meter; alternative to an
analogue quantity signal (0/4...20 mA).
The input signal stops any programmes running; the programmes will be
continued when the signal disappears.
The input signal triggers an "event sample". For example, the input
signal may be switched when a limit value is infringed; a separate bottle
can be filled in an event sample.
A message appears in the display showing the date and time (e.g.
function error of flowmeter). You must acknowledge the message; the
sample programme is not interrupted.
Special function for capacitive switch-off (optional)
Special function for capacitive switch-off (optional)
The input signal triggers a change to the switching programme.
6 Commissioning
79