Confirmation of error messages

Acknowledge error messages on the controller display by pressing the operating key
beneath the OK field.
5.5
Communication using ReadWin
Besides by means of local operation, you can also remotely configure the device and
poll measured values using a PC with the Endress+Hauser operating software
®
ReadWin
2000. ReadWin
operation. The interface for remote configuration is the RS232 connection socket. You
can find detailed information on operating the device using ReadWin
ROM with the device software supplied.
5.6
Calibration
5.6.1

Tap calibration

You must calibrate the tap if:
• The tap motor was replaced
• The error message: <Tap calibration> appears in the display.
Carry out the calibration as follows:
Fig. 20: Tap calibration
®
2000 is a universal PC programme for remote device
1. Switch the device on (see Section 6.2).
2. Under < SET -> SERVICE - CALIBRATION>, select
the item < DIST. TAB >.
3. When you select "Start", the tap turns and stops just
before the calibration position.
4. At the controller, keep selecting <1 step> until the
arrow on the front side of the tap is located exactly in
the notch in the middle of the distribution pan (point
A). In the picklist, select the menu option <SAVE>.
5. The tap is calibrated.