ASP station 2000
Endress+Hauser
Quantity
Ext. sig.
Flow
Shots
Distribution
Menu structure:
Program 1
Name
:
Program 1
Sample >
Distribution >
Start-Stop >
Synchronisation >
-mode
Parallel sample >
time
Esc
â
á
<-'
bottle
volume
Esc
Fig. 31: Distribution menu structure
You can programme a bottle change by time, number or an external signal:
Time
Number
Ext. sig.
Start-stop operation
Menu structure:
Program 1
Name
:
Program 1
Sample >
Distribution >
Start-Stop >
start
Synchronisation >
stop
Parallel sample >
operation
Esc
â
á
<-'
Fig. 32: Start-stop operation menu structure
You can start the sampling programme either immediately by pressing the AUT key, or
at a settable time. You can specify the sampling programme stop as follows:
Sampling is triggered after a specific measured flow rate.
!
Note!
For this function, a volume signal must be connected to the analogue input or
one of the digital inputs of the sampler. To do this, the digital input must be
configured using the <QUANTITY> function.
Sampling is triggered by an external signal.
!
Note!
One of the digital inputs must be assigned and configured using the
<SAMPLE> function.
Sampling is triggered at settable time intervals. The volume sampled is
proportional to a measured flow rate.
!
Note!
For this function, the ASP station 2000 must be equipped with a flow-
proportional dosing unit ("twiddle principle"). In addition, a volume signal
must be connected to the analogue input of the sampler.
Number of samples per sampling cycle.
Distribution
: time
time
:
00:15 number
:
12
ext.sig.
:
01,0 l
â
á
<-'
After a set time period has elapsed, the tap switches to the next empty bottle.
After a set number of samples has been taken, the tap switches to the next
empty bottle.
The tap switches to the next empty bottle when an external signal is present.
!
Note!
One digital input must be assigned and configured using the <BOTTLE>
function.
Start-Stop
: time
aut-button
:
prog.end
time
continuous
Esc
â
á
<-'
6 Commissioning
85