ASP station 2000
!
Endress+Hauser
Several programme combinations are available for the 7 programme version:
1
2
1+Event
2+Event
Main programmes
Four main programmes are available. Select a main programme from the item PROG.
NAME. (see "Creating a main programme" on page 84.)
Switching programmes
Two programme pairs (1⇔U1 and 2⇔U2) are available. The switching programmes (U1
and U2) are permanently assigned to the main programmes (1 and 2). You can trigger
a switch from the main programme to the switching programme as follows.
Day
Week
Q greater
Q smaller
Ext. signal
You can define separate sampling programmes and bottles for the switching
programmes.
Event programme
The event programme is activated by a digital input.
Note!
For this function, one of the digital inputs must be assigned and configured using the
<EVENT> function.
You can define a separate sampling programme and separate bottles for the event
programme.
A main programme is active.
A main programme and a switching programme are active.
A main programme and the event programme are active.
A main programme, a switching programme and the event programme are
active.
Switch to the switching programme at two settable times in one day.
Switch to the switching programme on three settable days in the week.
Switch to switching programme on exceeding a settable threshold value.
!
Note!
An analogue signal must be connected to the analogue input of the sampler
for this function.
Switch to switching programme on undershooting a settable threshold value.
Switch to switching programme by means of an external digital signal.
!
Note!
For this function, one of the digital inputs must be assigned and configured
using the <SWITCH> function.