Mobilize non-rigid components, such as set screws,
links, etc. during cleaning.
►
Thoroughly rinse through these components with
the cleaning disinfectant solution (at least five
times), using a disposable syringe.
Phase II
►
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
►
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
►
Drain any remaining water fully.
Phase III
►
Fully immerse the product in the disinfectant solu-
tion.
►
Mobilize non-rigid components, such as set screws,
joints, etc. during rinsing.
Manual cleaning with ultrasound and immersion disinfection
Phase
Step
I
Ultrasonic cleaning
II
Intermediate rinse
III
Disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water
quality at least)
RT:
Room temperature
*Recommended: BBraun Stabimed
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold)
>15
2
RT (cold)
1
-
RT (cold)
15
2
RT (cold)
1
-
RT
-
-
►
Rinse lumens at least 5 times at the beginning of
the exposure time using an appropriate disposable
syringe. Ensure that all accessible surfaces are
moistened.
Phase IV
►
Rinse/flush the product thoroughly (all accessible
surfaces).
►
Mobilize non-rigid components, such as set screws,
joints, etc. during final rinse.
►
Rinse lumens with an appropriate disposable
syringe at least five times.
►
Drain any remaining water fully.
Phase V
►
Dry the product in the drying phase with suitable
equipment (e.g. cloth, compressed air), see Vali-
dated cleaning and disinfection procedure.
Water
Chemical
quality
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
D–W
-
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
FD-W
-
-
-
13