Preparation before cleaning
Dismantle the product prior to cleaning, see Disas-
sembly.
Cleaning/disinfection
Product-specific safety notes on the repro-
cessing procedure
Damage to the product due to
inappropriate cleaning/disin-
fecting agents and/or exces-
CAUTION
sive temperatures!
To prevent increased contamination of loaded
instrument trays during use, please ensure that
contaminated instruments are collected separately
and not returned to the instrument tray.
Use suitable cleaning/disinfecting agents if the
product is put away in a wet condition. To prevent
foam formation and reduced effectiveness of the
process chemicals: Prior to mechanical cleaning
and disinfection, rinse the product thoroughly with
running water.
Use cleaning and disinfect-
ing agents according to the
manufacturer's instruc-
tions which
– are approved for use,
for example, on alumi-
num, plastic materials,
and high-grade steel,
– do not attack softeners
(e.g. in silicone).
Observe specifications
regarding concentration,
temperature, and exposure
time.
Carry out ultrasound cleaning:
– as an effective mechanical supplement to man-
ual cleaning/disinfecting.
– as a pre-cleaning procedure for products with
encrusted residues, in preparation for mechan-
ical cleaning/disinfecting.
– as an integrated mechanical support measure
for mechanical cleaning/disinfecting.
– for additional cleaning of products with resi-
dues left after mechanical cleaning/disinfect-
ing.
Clean and disinfect microsurgical products
mechanically if they can be placed securely in the
machine or on the positioning aids.