Instruments
Validated cleaning and disinfection procedure
Validated procedure
Manual cleaning with
immersion disinfection
only:
FW141P, FW143P,
FW228R, FW229R,
FW235R, FW238R,
FW239R, FW241R,
FW244P, FW258M,
FW262R–FW268R,
FW270R, FW271M,
FW272R, FW274M,
FW351R–FW355P
Manual cleaning with
ultrasound and immer-
sion disinfection
only:
FW177R, FW208R,
FW240R, FW242R
10
Specific requirements
Cleaning brush: TA011994/
TE654202, GK469200
20 ml disposable syringe
Keep working ends open for
cleaning.
When cleaning products with
movable hinges, ensure that these
are in an open position and, if
applicable, move the joint while
cleaning.
Drying phase: Use a lint-free cloth
or medical compressed air
Check visible surfaces for residues
after manual cleaning/disinfect-
ing.
Repeat the cleaning process if
necessary.
Cleaning brush: TA011994/
TE654202, GK469200
20 ml disposable syringe
Keep working ends open for
cleaning.
When cleaning products with
movable hinges, ensure that these
are in an open position and, if
applicable, move the joint while
cleaning.
Drying phase: Use a lint-free cloth
or medical compressed air
Reference
Chapter Manual cleaning/disinfection and
sub-chapter:
Chapter Manual cleaning with immer-
sion disinfection
Chapter Manual cleaning/disinfection and
sub-chapter:
Chapter Manual cleaning with ultra-
sound and immersion disinfection