10. Replacing fuses

Risk of death by electric shock!
Pull the mains plug before replacing the fuse
DANGER
sets.
Mandatory fuse set: 2x – T 1.25 AH, 250 V
Use a small screwdriver to release the clip on the fuse holder.
Pull out the fuse holder.
Replace both fuse sets.
Reinstall the fuse holder so that it audibly clicks into place.
Note
If the fuses burn out frequently, the device is faulty and should be repaired,
see Technical Service.

11. Technical Service

Danger to life of patients and users if the product
malfunctions and/or protective measures fail or are
not used!
DANGER
Do not modify the product.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
For
service
and
repairs,
B. Braun/Aesculapagency.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
8
please
contact
your
national

12. Accessories/Spare parts

Art. no.
Approvals
TE780
Europe
TE730
Europe
TE734
Great Britain
TE735
USA, Canada, Japan
Art. no.
Designation
TA020112
Fuse link T 1.25 AH, 250 V
TA014534
Instructions for use of Acculan 4 Charger GA320
(A4 for ring binder)

13. Technical data

13.1 Performance data, information about standards

Acculan 4 Charger GA320
Classification according to
Directive 93/42/EWG
Protection class according to
IEC/DIN EN 60601-1
Housing protection rating
according to
IEC/DIN EN 60529
For charging Aesculap bat-
tery types
Mains voltage ranges
(Current intake)
Frequency
Charging voltage/Output
voltage
Charging current/Output
current
Operational mode
Device protection
Weight without battery
Dimensions (L x W x H) with-
out battery
Dimensions (L x W x H) with
battery
Conforming to standard
EMC
CISPR11
Color
Length
[m]
Black
1.5
Black
5.0
Black
5.0
Gray
3.5
I
I
IP21
GA346 and GA666, GA676
100 V~ to 120 V~ (0.8 A to 0.7 A) /
220 V~ to 240 V~ (0.5 A to 0.4 A)
50 Hz to 60 Hz
max. 15 V
max. 1,2 A ± 0,3 A
Continuous operation
T 1.25 AH, 250 V
Form factor: 5 mm x 20 mm
approx. 2.3 kg
410 mm x 120 mm x 160 mm ±5 mm
410 mm x 252 mm x 160 mm ±5 mm
IEC/DIN EN 60601-1
IEC/DIN EN 60601-1-2
Class A

Werbung