The following table lists all logs and their content which the device outputs:
Log
Content
Weld log
The weld log is always output in the language that is currently chosen .
The weld log is a table with the following columns:
Date (dd . m m . y yyy)
Time (hh:mm:ss)
User (Name of the user who performed the welding)
Tube type (for example: C-Flex 374)
Tube description (for example: OD 1in / ID 3/4in)

Tube holder identification

Cycle (A counter that is incremented after every welding . )
Depyrogenation temperature
Welding temperature
Welding time (Cutting in to the tubes – tubes are pressed together)
Process time (Pressing start – Fan starting)
Event (Important event during the welding such as abort by user,
blocked condition, etc . )
Raw log
Please always send this file with a problem report/complaint .
Status
Please always send this file with a problem report/complaint .
9.8
Tube holder identification
The following list shows all currently supported tube holder identifications and their
tube geometries .
Tube holder
Tube geometry
identification |
Inside diameter
color code
[mm]
1
blue
19 .0
2
green
15 .9
3
grey
12 .7
4
white
9 .5
12
red
6 .4
9
orange
6 .4
11
yellow
3 .2
Outside diameter
[inches]
[mm]
3/4
25 .4
5/8
22 .2
1/2
19 .0
3/8
15 .9
1/4
11 .1
1/4
9 .5
1/8
6 .4
Appendix
[inches]
1
7/8
3/4
5/8
7/16
3/8
1/4
43