Herunterladen Inhalt Inhalt Diese Seite drucken

Sartorius Stedim Biotech BioWelder Total Containment TC 2000 Bedienungsanleitung Seite 46

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 47
Warning messages
There are no valid parameters on this
machine . Please load parameters .
SD missing . Please insert SD and retry!
SD not valid
Login failed
Insert validation holder
Remove validation holder
Run stopped by user
Log transfer stopped by user
46
Appendix
Cause
There are no parameters on the device .
No SD card inserted
Wrong SD card inserted
Wrong user name or password entered
1 . Validation holder not inserted
2 . Validation holder inserted, identification
defective
1 . Validation holder was not removed
2 . Validation holder was removed, identifica-
tion defective
User interrupts a sequence
Abort by the user while data being written to
the SD card
Elimination
Switch machine on and off .
If there is still no current data,
contact your Sartorius Stedim
Biotech Service Representative .
Insert card in SD slot and repeat the
action again .
Use the SD card supplied by Sartorius
Stedim Biotech .
Enter correct user name with the
associated password
1 . Insert validation holder
2 . Contact Sartorius Stedim Biotech
Service Representative .
1 . Remove validation holder
2 . Contact your Sartorius Stedim
Biotech Service Representative .
Confirm message with OK
Confirm message with OK

Werbung