Before start-up check whether the rated voltage of the device coincides with
the local line voltage.
The device was developed for laboratory use and must be protected from every
type of fluid or vapour.
The device can be operated between +5°C and +40°C.
The welding parameters are qualified for room temperature (20°C – 25°C)
operation only.
Optimal relative humidity is 60%.
®
The device is programmed with standard parameter sets for welding C-Flex
374,
®
®
AdvantaFlex
, PharMed
BPT and SaniPure™ BDF™.
These parameter sets have been validated at ambiant temperature.
The degree of protection is IP20.
Make sure that the device is not switched off during the welding process.
Ensure that there is an uninterrupted power supply (UPS).
Use only the blades provided by the manufacturer and use it only once.
Use only tested control cables and accessories.
Isolate the supply voltage before replacing a fuse.
The device has an Ethernet connection. Only skilled personnel may connect the
device to a network.
Only shielded Ethernet cables may be used.
Only skilled personnel may configure the device for connection to a network.
It is recommended to operate the device via the touch screen during sequences.
The device contains electric drives which can be blocked | destroyed by
foreign substances. Make sure that no foreign substances reach the interior
of the device.
®
After using the BioWelder
TC, you should always log out. Otherwise another
user is able to manipulate the device with your account.
®
The auto login is used to work more efficiently with the BioWelder
TC.
The auto login can only be activated for one user. This user will be logged in
AUTOMATICALLY without entering any username and password after each
restart or after changing the user interface.
A user of the Supervisor group should NOT use the auto login because
of safety reasons, otherwise any user is able to take essential changes on
®
the BioWelder
TC under the Supervisor account.
Not following the manual instructions of the manufacturer may lead in to
unreliability, hazardous situations and disturbances of safety.
6