Guidance and manufacturer's declaration – electromagnetic immunity
XENON 100 model 20 1325 20 is intended for use in the electromagnetic environment specified below. The customer or user of the XENON 100 model
20 1325 20 should ensure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
IEC 61000-4-2
± 8 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
± 1 kV differential mode
IEC 61000-4-5
± 2 kV common mode
Voltage dips, short
interruptions and voltage
<5% U
variations on power supply
(>95% dip in U
input lines
for 0.5 cycle
IEC 61000-4-11
40% U
(60% dip in U
for 5 cycles
70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
Appendix
Electromagnetic Compatibility (EMC)
Information
Table 202
Compliance level
Complies
± 6 kV contact
± 8 kV air
Complies
± 2 kV for power supply
lines
± 1 kV for input/output lines
Complies
± 1 kV differential mode
± 2 kV common mode
Complies
*
<5% U
*
T
T
)
(>95% dip in U
)
T
T
for 0.5 cycle
Complies
40% U
T
T
)
(60% dip in U
)
T
T
for 5 cycles
Complies
70% U
T
T
)
(30% dip in U
)
T
T
for 25 cycles
T
)
See guidelines
T
Complies
3 A/m
Ïðèëîæåíèå
Óêàçàíèÿ ïî ýëåêòðîìàãíèòíîé
ñîâìåñòèìîñòè (ÝÌÑ)
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
If the user of the equipment or system requires
continued operation during power mains interruptions,
or operation at <5% U
(>95% dip in U
) for 5 seconds,
T
T
it is recommended that the equipment or system be
powered from an uninterruptible power supply or a
battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.