 When applying the lockable castors (B), make sure that Tropic
plus is positioned between the lockable castors.
Preparation

Connecting to power supply

WARNING
Risk of electric shock
Connecting devices to auxiliary power sockets may
increase leakage current. If the protective ground on one of
these devices fails, the leakage current may exceed the per-
missible values. Only make connections after agreement
with the respective manufacturer of the device. Have ser-
vice personnel check the leakage current.
If the permissible value is exceeded, do not use the auxilia-
ry power sockets on the device but separate mains power
sockets on the wall.
CAUTION
Device inoperable
Before putting the device into operation for the first time, the
device voltage must be set to the mains voltage available using
the voltage switch at the back. 230 V is preset.
Select mains voltage:
 Carefully unlock the fuse insert (A) on the device socket using
a screwdriver and remove it.
 Remove the voltage switch (B) from the fuse insert and select
the required mains voltage by rotating it. Put the fuse insert
back in by pressing it until it audibly clicks into place.
22
B
B
B
C
A
 The required mains voltage (110 V/127 V/230 V) is now visible
in the window (C).
NOTE
Observe the rating plate on the underside of the device.
E
 Connect the power cable (D) to the device power socket and
to a grounded mains power socket.
Use with heated hose
Use only original Dräger hose. Malfunctions may arise if other
heated hoses are used.
 Connect the plug of the hose heater (E) to the socket at the
back of the device. Make sure the mark on the plug is pointing
upwards.
 Attach the heated silicone hose (instead of the long 1000 mm
hose) onto the opening of the nebulizer chamber lid, the refill-
ing system, or the medication container.
The hose heater switches on automatically when the ultrasonic
nebulizer starts. The operating temperature of max. 40 °C is
reached after approx. 10 minutes.
CAUTION
Risk of patient injury
Only use heated hoses up to an ambient temperature of 30 °C.