Herunterladen Inhalt Inhalt Diese Seite drucken

Information For Use - Ottobock 3R67 Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 12
8 Use

8.1 Information for use

NOTICE
Mechanical overload
Impaired functionality due to mechanical damage
Check the product for damage prior to each use.
Do not use the product if its functionality has been impaired.
Take any necessary measures (e.g. repair, replacement, inspection by the manufacturer's
customer service, etc.).
8.2 Cleaning
CAUTION
Use of unsuitable cleaning agents or disinfectants
Impairment of functionality and damage due to incorrect cleaning agents or disinfectants
Clean the product only according to the instructions given in this section.
For the prosthesis, observe the cleaning instructions for all prosthetic components.
Cleaning light soiling
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
3) Allow to air dry in order to remove residual moisture.
Cleaning heavier soiling
Required materials: cleaning cloth, 634A58 isopropyl alcohol, soft cloth
>
► NOTICE! Verify material compatibility. Do not remove lubricant from bearing points.
► Clean the product with a cleaning cloth and isopropyl alcohol.
► Dry the product with a cloth.
Cleaning with Disinfectants
Required materials: colourless, alcohol-free disinfectant (verify material compatibility!), soft
>
cloth
► Disinfect the product with the disinfectant.
► Dry the product with a cloth.
► Allow to air dry in order to remove residual moisture.
9 Maintenance
CAUTION
Failure to follow the maintenance instructions
Risk of injuries due to changes in or loss of functionality and damage to the product
Observe the following maintenance instructions.
► NOTICE! Do not lubricate and grease the prosthetic joint.
► NOTICE! Repair work must be performed exclusively by the manufacturer's technical service.
► Arrange regular maintenance intervals with the patient depending on the level of use.
► Following an individual period for the patient to get accustomed to the prosthesis, check the
settings of the prosthetic joint and adapt them to the patient's requirements again as needed.
► The prosthetic components should be inspected after the first 30 days of use.
► Inspect the entire prosthesis for wear during normal consultations.
► Perform a safety inspection every 3 months.
25

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis