2 Intended use

2.1 Indications for use
The product is intended exclusively for use in pairs, in orthotic fit­
tings of the lower limbs and exclusively for contact with intact skin.
The orthosis must be used in accordance with the indications.
2.2 Indications
Dorsiflexor weakness
Indications must be determined by the physician.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or
CAUTION
injury.
Warning regarding possible technical damage.
NOTICE
3.2 General safety instructions
CAUTION
Incorrect alignment, assembly or adjustment
Injury and damage to product components
► Observe the alignment, assembly and adjustment instructions.
► Work on the product may only be carried out by qualified person­
nel.
CAUTION
Use of the product in case of known allergies to chrome or
nickel
Allergic reaction
► Avoid skin contact with the product in case of known allergies to
chrome or nickel.
CAUTION
Mechanical damage to the product
Injuries due to changes in or loss of functionality
► Use caution when working with the product.
► Check the product for proper function and readiness for use.
► In case of changes in or loss of functionality, discontinue use of
the product and have it checked by authorised, qualified person­
nel.

4 Preparation for use

4.1 Environmental conditions
Permitted: Temperature range for use −10 °C to +60 °C; relative
humidity 0 % to 90 %, non-condensing.
Not permitted: Mechanical vibrations, impacts, perspiration, urine,
fresh water, salt water, acids, dust, sand, highly hygroscopic particles
(such as talcum).
4.2 Processing
For further processing instructions, see the 646T523=* technical
information.
1) Place the dummy on the joint pivot point of the plaster model and
secure it with the nails (see fig. 1).
2) Vacuum-form or laminate the orthosis.
3) Remove the orthosis and dummies from the plaster model after it
has hardened/cooled.
4) Separate the foot and calf component with a saw and deburr the
edges (see fig. 2).
5) For pre-flexed joints make a U-shaped cutout, for straight joints
make a V-shaped cutout at the front sides of the joint adapters
(see fig. 3).
6) Drill the mounting holes for the joints (diameter 4.5 mm).
7) Insert the joints and attach them to the orthosis with the screws
(see fig. 4) (torque value: 0.5 Nm).
8) Cover both joints with the padding.