5 Preparing the product for use

CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
► Observe the alignment and assembly instructions.
INFORMATION
► Not all of the materials may be available in your country. In this
case, please contact the local branch of the manufacturer to
obtain information on alternative materials.
5.1 Installing the foot adapter
► Install the foot adapter on the prosthetic foot according to its

instructions for use.

12
ery)
Reference num­
ber
2R8*, 2R31*,
2R54*
2R14
2K34=*
2Z22=M10
636W28 or
636W18 and
636W19
5.2 Installing the ankle block (exoskeletal design)
Required materials: Screw connection (see Page 11), 710D4
>
torque wrench, sandpaper (grain size 40), 636W28 Ottobock
special adhesive and hardener, suitable solvent (e.g. 634A28)
1) Insert the threaded socket of the screw connection in the proximal
side of the hole in the ankle block.
2) Insert the screw on the foot sole side in the prosthetic foot and
screw the ankle block to the prosthetic foot (torque: 30 Nm).
3) Conduct the fitting of the ankle block.
4) Loosen the screw connection between the prosthetic foot and the
ankle block.
5) Sand the ankle block according to the contour of the prosthetic
foot.
6) Roughen the connection surface of the prosthetic foot in an a – p
direction using the sandpaper.
7) Remove the dust from the sanded surfaces.
8) Mix the adhesive for the connection between the ankle block and
the prosthetic foot.
9) INFORMATION: Do not put adhesive in the hole for the
screw connection.
Apply adhesive evenly to the connection surfaces of the ankle
block and prosthetic foot.
10) Align the ankle block and prosthetic foot with each other.
11) Insert the screw on the foot sole side in the prosthetic foot and
screw the ankle block to the prosthetic foot (torque: 30 Nm).
12) Wipe off adhesive that comes out of the joint using a cloth.
13) Remove any remnants of the adhesive using a suitable solvent.
14) Allow the adhesive to cure (curing time: 16 h).