Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker Precision AC Bedienungsanleitung Seite 13

20 mm coupler, c-mount
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 33
Rinse 1
Recirculation Time
Water Temperature
Detergent Type
Dry Phase
Recirculation Time
Water Temperature
Detergent Type
• Filtered pressurized air (≤40 psi) can be used to aid in drying.
• Visually inspect the device for cleanliness, paying close
attention to hard-to-reach areas. If visible soil remains, repeat
steps 1 through 4.
Drying
• For automated drying, use the drying cycle provided with the
washer.
• For manual drying, use a clean, lint-free cloth.
• Filtered pressurized air (≤40 psi) can be used to aid in drying any
crevices, mated surfaces, and hard-to-reach areas.
Disinfection (Optional)
Thermal Disinfection
The device must be sterilized after disinfection. Failure
to sterilize the device before reuse presents an acute
infection control risk to the patient.
Follow the steps below to disinfect the device using thermal
disinfection.
1. Place the device into the washer disinfector.
2. Program the washer using the following parameters:
Thermal Disinfection
Recirculation Time
Water Temperature
2 Minutes
Hot Tap Water
N/A
7 Minutes
115°C
N/A
1 Minute
90°C
EN-9

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis