Instructions for Reprocessing
Point of Use
• Disassemble the coupler from the scope and camera head.
To disconnect the scope, depress the endobody clamp on the
coupler and remove the scope from the coupler. To disconnect
the coupler, grip the rear adapter of the coupler and unscrew it
counterclockwise from the camera head.
• Wipe excess soil from the device with a clean sterile cloth.
Containment and Transportation
• Reprocess the device as soon as reasonably practical following
use.
• Transport the device in a tray to avoid damage. Follow
the facility's internal procedures for the transportation of
contaminated surgical instruments and devices.
Cleaning
Manual Cleaning
1. Soak
• Prepare a solution of enzymatic detergent and tap water at
35°C. Refer to the manufacturer's instructions for detergent
concentration and any other recommendations.
• Fully immerse the device and use an appropriately-sized
syringe to inject any mated surfaces and crevices with at least
50 mL of the prepared detergent.
• Soak the device for a minimum of 15 minutes.
2. Brush
• Thoroughly brush the exterior of the device with a soft-
bristled brush for 1 minute, focusing on any crevices and
mated surfaces.
• Inject any mated surfaces and crevices a minimum of 5 times
with 50 mL of the prepared detergent.
• While pushing down on the endobody clamp, brush between
all gaps and crevices. Continue brushing all gaps and crevices
while releasing the clamp to the initial position. (Refer to the
"Operating the Endobody Clamp" section for more detail
about moving the clamp.)
3. Rinse
• Remove the device from the prepared detergent. Rinse the
device with reverse osmosis/deionized water at ambient
temperature for 1 minute or until all visible detergent residue
has been removed.
• Flush any mated surfaces and crevices a minimum of 5 times.
After all visible detergent residue has been removed, continue
to rinse for 30 seconds.
• Drain excess water from the device and dry it using a clean
cloth or filtered pressurized air.
• Visually inspect the device for cleanliness, paying close
attention to hard-to-reach areas. If visible soil remains, repeat
steps 1 through 3.