English
Precision AC 20 mm Coupler, C-Mount
Product Description and Intended Use
The Stryker Precision AC Coupler is used to attach endoscopes to
Stryker surgical camera heads that do not have an integrated coupler.
The coupler facilitates the easy exchange of different-sized scopes
during a surgical procedure and allows users to focus the image
captured through the scope.
1
2
1. Focusing ring
2. Rear adapter (camera end)
3. Endobody clamp
4. Endobody (scope end) - marked "AUTOCLAVE" on the face of the
endobody to indicate compatibility with steam sterilization
Indications
The coupler is indicated for use in minimally invasive surgical
procedures where a Stryker endoscopic camera and endoscope are
used.
Contraindications
There are no known contraindications.
Warnings and Cautions
1. Read these instructions thoroughly before using the device.
2. Before using this device, read the Precision AC Video Camera user
guide (Stryker part P26182 for English, or P26185 for multilingual)
for warnings, indications, intended use, and other information
about using the camera system.
3. Federal law (USA) restricts this device to use by, or on order of, a
physician.
4. Inspect the coupler for any damage that might have occurred
during shipment or prior use.
5. The coupler contains optomechanical parts that are precisely
aligned. Dropping the coupler may cause mechanical damage and/
or misalignment.
6. Test the coupler for functionality before using it in a surgical
procedure.
7. This coupler is shipped non-sterile. Clean and sterilize the coupler
prior to the first use and after every subsequent use. Follow the
cleaning and sterilization instructions provided in this user guide.
8. No modification of this device is authorized by Stryker.
3
4
"AUTOCLAVE"
EN-1