Steris/Amsco V-PRO
1. Clean and prepare the coupler as recommended in this user
guide.
2. If using a sterilization tray (optional), follow any additional
instructions provided with the tray. Use only trays that are
approved for sterilization with Steris/Amsco V-PRO.
3. Double wrap the coupler (or tray) prior to sterilization.
4. Sterilize the device using one of the following V-PRO sterilization
systems:
• V-PRO maX Sterilizer (Non-Lumen or Standard cycle)
• V-PRO 1 Plus Sterilizer (Non-Lumen or Standard cycle)
• V-PRO 1 Sterilizer (Standard cycle).
STERRAD®
1. Clean and prepare the coupler as recommended in this user
guide.
2. If using a sterilization tray (optional), follow any additional
instructions provided with the tray. Use only trays that are
compatible with Sterrad.
Not all sterilization trays are compatible with STERRAD®
systems. Using an incompatible tray may result in
incomplete device sterilization. Consult the instructions
that came with your sterilization tray to determine which
sterilization method is compatible with your tray and
devices. If a compatible tray is not available, the devices
can be double-wrapped for use with the STERRAD®
system.
3. Double wrap the coupler (or tray) prior to sterilization.
4. Sterilize the coupler following the instructions of the
manufacturer, using the Sterrad 100S, 200, NX, or 100NX
Sterilization System. Select the standard cycle.
EN-11