2. Depress the endobody clamp
endobody
Before each use, check the outer surface of the
endoscope to ensure there are no rough surfaces, sharp
edges, or protrusions.
3. Release the endobody clamp to secure the scope.
4. Attach a light cable from the light source to the light post on the
endoscope
5. With the camera console turned on, rotate the coupler's focusing
ring
to establish a sharp focus on the display monitor.
4
Note: To remove the coupler, grip the rear adapter and unscrew the
coupler counterclockwise from the camera head.
Reprocessing
These reprocessing instructions are provided in accordance with
ISO 17664, ISO 17665, AAMI TIR12, AAMI ST79, and AAMI ST81.
The instructions have been validated by Stryker as being capable
of preparing the device for re-use. To achieve the desired result,
the processor shall ensure that the following instructions are
performed as written in their entirety and as appropriate in the
processor's facility. This normally requires routine monitoring
and validation of the facility's reprocessing procedures. Stryker
recommends users observe these standards when reprocessing
medical devices.
Overview
Reprocessing the coupler involves manual or automated
cleaning with two different detergents, optional disinfection, and
sterilization.
• Step 1 (required): Cleaning with Enzymatic Detergent
• Step 2 (required): Cleaning with Non-Enzymatic Detergent
• Step 3 (optional): Disinfection
• Step 4 (required): Sterilization
.
2
1
4
.
3
and insert a scope into the
1
2
3
EN-3