11.2 Quantitative interpretation from plasma (EDTA)
To quantify CMV-positive specimens, a standard curve measurement using standard
A, B, C, and D must be performed.
Fig. 2:
Standard CMV series (10
11.2.1 Cycler linearity
®
The RIDA
GENE CMV real-time PCR test is linear in a range of 10
1,000,000) genome equivalents per reaction (gEq/reaction) for all real-time PCR
devices described here.
Table 8: Measurement range of all real-time PCR devices described here
Result specimen for CMV/reaction
Concentration* > 1x10
2
1x10
≤ concentration* ≤ 1x10
Concentration* < 1x10
*Designation can deviate depending on the device.
Note: For further information on specimen quantification, please contact
pcr@r-biopharm.de
RIDA
GENE CMV
2
5
to 10
gEq/reaction) on LightCycler
6
6
2
.
2019-05-28
Interpretation/indication of results
(gEq/reaction)
More than 1,000,000
= Quantity
Less than 100
®
480II
2
6
to 10
(100 to
35