Table 6: A valid quantitative PCR run must meet the following conditions:
CMV standard curve
Correlation coefficient
5
Standard D (10
)
PTC
NTC
If the positive control (PTC) is not detected in the specified Ct range but the negative
control is valid, all reactions need to be reanalyzed, including the positive control.
If the negative control (NTC) is not negative but the positive control is valid, all
reactions need to be reanalyzed, including the negative control.
If the specified values are not met, check the following things before repeating the test:
- Expiration date of the reagents used
- Functionality of the devices used
- Correct test procedure
Fig. 1: Correct run of the positive and negative control (CMV) on LightCycler
RIDA
GENE CMV
Quantitative analysis
Result
Positive
Positive
Positive
Negative
2019-05-28
Acceptance range
2
R
2
0.990 ≤ R
≤ 1.000
Ct
Ct ≤ 25
See Quality
Assurance
Certificate
0
®
480II
33