Transporting a patient with the Big Wheel

EN
WARNING
Always position the patient in the center of the product.
Always lock the siderails in the full up position with the sleep surface horizontal and in the lowest position when you
transport a patient.
CAUTION
Do not apply the steer pedal when the Big Wheel is resting on a threshold or other raised area. The force required to
apply the Big Wheel will be higher than normal which may cause product damage.
To transport a patient with the Big Wheel:
1.
Raise the siderails to the highest position. Pull on the siderail to make sure that it is latched.
2.
Lower the litter to the lowest position.
3.
For Model 1125 only, make sure that the On/Drive-Off/Manual switch is in the Off position.
4.
Unplug the power cord from the wall outlet if your product is equipped with the electric lift or electric litter option.
5.
Put the brake/steer pedal (A) in the steer position
6.
Use the push handles to push the product.
7.
Apply the brakes to lock the product in place.
Notes
Always make sure that the brake is released before you attempt to move the product. Attempting to move the
product with the brakes applied may result in operator or patient injury.
The Big Wheel (B) does not pivot. You cannot move the product sideways with the Big Wheel applied.
When the Big Wheel is applied, the product is raised and the foot end casters are elevated. The product rests on
the head end casters and the Big Wheel to make the product easier to steer.

Transporting a patient with the Zoom drive wheel

WARNING
Always position the patient in the center of the product.
Always lock the siderails in the full up position with the sleep surface horizontal and in the lowest position when you
transport a patient.
Do not use the Prime Series stretcher with the Zoom Motorized Drive in the MR environment. The Prime Series
stretcher with the Zoom Motorized Drive is MR (Magnetic Resonance Imaging) Unsafe.
Always operate the product when all operators are clear of the mechanisms.
CAUTION
Always immediately unplug the power cord from the wall outlet and turn the On/Drive-Off/Manual switch to the Off
position if large fluid spills occur in the area of the circuit boards or motors. Fluids can short out controls and may cause
the product to operate erratically or may make some functions inoperable. Component failure caused by fluids may also
cause the product to operate unpredictably and could cause injury to the patient. Do not put the product back into
service until it is dry and has been tested for safe operation.
To transport a patient with the Zoom Motorized Drive:
1.
Unplug the power cord from the wall outlet.
1-26
Operation
(Figure 1-7 on page
1-25).
1125-309-005 REV A
www.stryker.com