Locating the patient restraint strap tie-ins (Continued)
Figure 1-22: Prime restraint strap locations
Note: Restraint straps are Type B applied parts.

Positioning the upright X-ray cassette holder - Prime X option

WARNING
Always take protective measures when you use the upright or lateral X-ray cassette holder. The X-ray cassette
holder does not protect against radiation.
Always refer to local, state, and federal guidelines in addition to facility protocols for safety before you use Prime X
option with radiation generating devices. Radiation generating devices may produce residual, stray, or scattered
radiation.
Always use extra caution when you use a mattress thicker than 2.5 in. or a mattress overlay with Prime X option.
Always follow the Positioning the upright X-ray cassette holder - Prime X option instructions to insert the X-ray
cassette.
The X-ray cassette holder attaches to the Fowler backrest weldment to support X-ray cassettes. You can take X-rays
while a patient is on the product. You can also adjust the cassette's position before you take an X-ray.
To position the X-ray cassette holder:
1.
Apply the brakes. Push on the product to make sure that the brakes are working.
2.
Raise the Fowler backrest section to full height.
3.
Insert the lower retainer guides (A) under the Fowler backrest weldment bar
4.
Raise the cassette holder until the retainer guides latch onto the Fowler backrest weldment. Make sure that the
cassette holder is secure.
www.stryker.com
Accessories
Figure 1-23: Prime X restraint strap locations
1125-309-005 REV A
(Figure 1-24 on page
1-50).
English
EN
1-49