The ratio of glycated hemoglobin to total hemoglobin is determined
and the result is presented in up to two user selectable units: %
DCCT, mmol/mol IFCC, % JDS, eAG mg/dl or eAG mmol/l.
Linear relationships have been established between these reportable
units
(4,5)
:
mmol/mol IFCC = (% DCCT – 2.15) x 10.929
% JDS = (0.09274 x mmol/mol IFCC) + 1.724
eAG mg/dl = (28.7 x % DCCT) - 46.7
eAG mmol/l = (1.59 x % DCCT) - 2.59
eAG values are based on a correlation study linking % DCCT to the
patient's average blood glucose concentration, resulting in published
formula to derive the eAG
(5)
.
Cartridge Pack Contents
15 Quo-Test A1C Test Cartridges packed individually into foil
pouches an indicating silica gel sachet.
A minimum of 15 Quo-Test Blood Collectors.
1 Calibration Barcode (on inner flap of carton, indicated by the CAL
symbol).
1 Instruction For Use leaflet.
Accessories Required (not included)
Quo-Test Analyzer
Quo-Test A1C Control Kit
Standard Blood Collection Equipment
Active Ingredients
Fluorescent boronate conjugate
Plastic vial containing ammonium chloride buffer, lysis agent and
preservative
Blood Collector treated with EDTA and surfactant
Calibration of the Quo-Test A1C Test
The Quo-Test A1C System is certified by the NGSP and IFCC
(6)
.
The Quo- Test Analyzer and A1C Test Cartridges have been
calibrated using samples provided by the ERL via the NGSP network.
Results obtained using the Quo-Test A1C System are traceable to
the IFCC reference method
(7)
.