How to Run a Quo-Test A1C Test
Refer to images 1 - 9 on the inside of the back cover.
Step 1
Ensure that the analyzer is ready to run a test. Refer to the Quo-Test
Analyzer System User Manual for details.
Step 2
Before using a new lot of Test Cartridges, scan the calibration
barcode that is printed on the inner flap of the carton, as indicated by
the CAL symbol.
Place the barcode scanner over the calibration barcode and press
the barcode scanner button until you hear a beep. The analyzer will
confirm that the data has been accepted, along with displaying the lot
number and expiry date.
Step 3
Remove a pouched cartridge from the carton and place it next to the
analyzer. Allow at least 40 minutes for it to equilibrate to the
analyzer's environmental temperature before use, even if previously
stored at room temperature.
When ready to perform the analysis, carefully open the pouch and
remove the Test Cartridge and silica gel sachet. Check the indicator
colour of the silica gel (see colour chart inside back cover).
Do not handle the lower part of the cartridge containing the liquid.
Do not use the cartridge if condensation is present.
After opening the pouch, the cartridge must be used within
5 minutes.
Step 4
For finger prick sample collection, the patient's finger must be warm,
dry and clean, including being free of substances such as hand
cream. Use a single use lancet (not provided) on the patient's finger
to get a droplet of blood about the same width as the Blood Collector.
For venous samples previously collected into EDTA tubes, ensure
that the sample is thoroughly mixed and use standard sampling
equipment to place a droplet of blood, approximately 10 μl to 20 μl,
onto a non-metallic and non-absorbent surface such as
Parafilm
®
M
(9)
. Do not attempt to sample directly from the tube.
Step 5
Lightly touch the pointed tip of the Blood Collector to the top surface
of the droplet of blood, as per the image. The blood will be taken up
by the collector. The blood must completely fill the slot.
Note: Avoid taking up air bubbles and excess sample on the outside
of the tip of the Blood Collector. Do not wipe excess blood from the
collector.
9
EN
3121-9101-0255_ML1_02.01_QT Test Kit - 05/2017