Step 6
Cartridge, making sure that the top of the Blood Collector is flush with
the top of the Test Cartridge and not sticking out. Do not push the
Blood Collector into the Test Cartridge, as this may cause an error
message to be displayed instead of a result. Refer to the Quo-Test
Analyzer System User Manual for further details.
Step 7
After the Test Cartridge has been prepared, open the door of the
analyzer and place the Test Cartridge into the test chamber, Making
sure that the Test Cartridge is firmly seated.
The Test Cartridge must be inserted into the analyzer within 1 minute
of the blood sample being collected.
Step 8
Pull the red slide towards the front of the analyzer and then close the
door. The test will start automatically. You will be prompted to scan
the Operator and Patient ID if these features have been enabled.
Note: If the analyzer displays 'Scan Lot ****** Data Card' then scan
the calibration barcode (see Step 2).
Step 9
When the test has finished the result will be displayed on the screen.
Open the door and push the red slide towards the rear of the
analyzer. Remove the Test Cartridge and dispose of it as clinical
waste. Upon closing the analyzer door, the display will return to the
home screen and the analyzer will be ready to run another sample.
Note: Refer to the Quo-Test Analyzer User Manual to access
previous results in memory.