14.3.1 Cleaning the lenses
SI Analytics® advises to regularly check that the 2 lenses inside the cell holder are clean.
In order to access the lenses, use the Quick Lock system to take out the accessory inserted
in the cell holder (see pictures). If necessary, then proceed to the cautious cleaning of the
2 lenses using a cotton bud lightly soaked with isopropanol.
Check there are no deposit/traces before use.
Quick Lock

14.4 What to do, if ...?

14.4.1 Measures if cell is broken
WARNING
Cells can contain dangerous substances. If the contents are released, follow the
safety instructions in the package insert. If necessary, take corresponding
protective measures (protective goggles, protective gloves etc.).
CAUTION
Do not turn the photometer upside down to remove the liquid! When doing so, the
liquid could come into contact with electronic components and damage the
photometer.
The photometer has a drain device through which the contents of a broken cell can drain
off without causing any damage.
Proceeding after a cuvette is broken.
1. Switch off the spectrophotometer and disconnect it from the main power supply.
2. Let the liquid drain off into a suitable container and dispose of it properly according to the
instructions of the reagent package.
3. Carefully clean the cuvette well using a wet, lint-free cloth.
4. Let dry the cuvette well.
Remark
After recommissioning, check out the spectrophotometer for all measurements.
Maintenance
Lenses