Sl Analytics UviLine Sl 5000 Gebrauchsanleitung Seite 131

Spectrophotometer
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Start up
133
1. With « ▲ » and « ▼ » keys, select one option from the menu. Confirm this option with
<START·ENTER> key.
The input field for the current date pops up.
2. Enter the current date with <0...9> keys and confirm.
The input field closes. The date is accepted.
3. Select « Time » and confirm.
The input field for the current hour pops up.
4. Enter the current hour with <0...9> keys and confirm.
The input field closes. The time is accepted.
NOTE: After the initial commissioning, the operator can change the date and time at any time.
From the main menu <HOME> enter in the menu "General setup", then "Date/time" menu.
During the initial commissioning, the instrument automatically guides the operator to the self-test
after the setting of the date and time.
5.2.6 Self test
IMPORTANT
During self-test, the cuvette compartment must be empty (without any cuvette inside the cuvette
well) and the cover of the cuvette compartment must be closed.
1. Be sure that the cuvette compartment is empty.
2. Press <START ENTER> key to run the self-test.