- Press F1 key [New method]. It displays:
The following parameters can be modfied:
* Number
Automatic classification of methods (From 1001 to 1100).
* Wavelength
320 – 1100 nm for UviLine
190 – 1100 nm for UviLine
* Resolution
From 0 to 3
It is the number of figures after comma for the result.
* Unit
i.e.: mg/L (10 characters maximum).
* Citation form
(18 characters maximum).
Enter the chemical formula of the compound to be measured. i.e.: NO
* Designation
(18 characters maximum).
Enter the compound name. i.e.: Nitrate
* Version
(18 characters maximum).
Enter the parameter concentration range. i.e.: 0.9 – 75.3 mg/L
Concentration mode
SI
5000/6000
SI
7000