10. Specifications
10. Specifications
10.1 Specifications for CPC2000
Reference pressure sensor
Pressure range
Gauge pressure
1)
Accuracy
Calibration interval
1) It is defined by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following
factors: the intrinsic performance of the instrument, the measurement uncertainty of the reference instrument, long-term
stability, influence of ambient conditions, drift and temperature effects over the compensated range during a periodic zero point
correction.
2) FS = Full scale = End of measuring range - Start of measuring range
Pressure controller
Instrument
Instrument version
Weight
Warm-up time
Ingress protection
Digital display
Type of display
Resolution
Keyboard
Menu languages
Measuring range
Pressure type
Functions
Units
Zero point adjustment
Pressure generation
WIKA operating instructions, model CPC2000
0 ... 1 mbar
2)
0.3 % FS
365 days
Desktop instrument with carrying handle
4.6 kg [10.14 lb]
Approx. 15 min
IP20
Backlit, multiline graphic display
4 ... 5 digits depending on range and units
Membrane keyboard
German
■
English
■
Spanish
■
French
■
1 ... 1,000 mbar
Gauge
Pa
hPa
■
■
kPa
bar
■
■
Automatic (at definable time intervals)
■
Manual (ZERO button)
■
Internal, electric pump
0 ... 10 mbar
■
0 ... 50 mbar
■
0 ... 100 mbar
■
0 ... 500 mbar
■
0 ... 1.000 mbar
■
2)
0.1 % FS
mbar
inH
O
■
■
2
psi
inHg
■
■
EN
mmHg
■
Torr
■
39