WIKA Mensor CPC2000 Betriebsanleitung Seite 24

Vorschau ausblenden Andere Handbücher für Mensor CPC2000:
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

5. Commissioning and operation
To exit the submenu, press the [MENU] button.
Usage
On leaving the menu, the controller will reverter to standby status. All information about
the automatic mode will be shown on the display.
EN
VENT
6
Cycl.
6
1
/ 10
5
Start
1
2
4
5
Next cycle
End of cycle
The figure shows the process description for the automatic mode. Based on the table,
one will know which cycles are running and the meaning of each.
Phase
Meaning
1
Wait for the start command with [SELECT]
Delay time that can be defined in the configuration
2
4
Duration of the system zero point adjustment
7
Hold time at the zero point
9
Time to adjust to the next step
10
Hold time
13
Delay time at the maximum target pressure
15
Time to adjust to the next step
16
Delay time at the zero point
24
1
+10.000
mbar
Phase
1
Time
5 s
Press
-0.007
+
0%
+0.000
4
7
9 10 9 10 9 10 9 10 13 15 16 15 16 15 16 15 16 18
End value
1
Pressure unit
2
2
Phase 1: it is being vented
3
(Pressure = 0)
3
Display of the current target
4
pressure in percent
Battery status
5
First of ten cycles
6
Operating mode
6
WIKA operating instructions, model CPC2000