5. Commissioning and operation
5.8
SETUP settings (SETUP)
The SETUP menu items are made up of six submenu items.
EN
Menu item
RANGE
UNIT
STEPS
MODE
LANGUAGE
SETTINGS
26
MENU
-------------------------
RANGE.....:
UNIT......:
STEPS.....:
MODE......:
LANGUAGE..:
SETTINGS
Meaning
Setting of the pressure range (max. 100 % of the measuring span)
Selection of the pressure unit
Selectable pressure units are:
Pa
hPa
■
■
kPa
bar
■
■
The selected pressure unit will be shown on the display.
Dependent upon the measuring range of the instrument, various units
are not available since they could not be displayed.
Setting of the step size in x %.
The step size for the STEP function, in x % of the set range, can be
changed.
Possible settings are from 1 ... 50 %.
Selection and setting of the operating mode
The following modes can be selected:
Measure "MEAS"
Control "CTRL"
Automatic "AUTO"
Selection of the display language
The language in which the menu will be displayed can be set here.
Following language can be selected:
German
■
English
■
Here additional instrument settings can be made
+10.000
mbar
50%
MEAS
EN
-->
mbar
inH₂O
■
■
psi
inHg
■
■
Measurement of differential and gauge
pressure in the set pressure unit
Regulates to the specified set point
automatically
Carries out a stored test routine
Spanish
■
French
■
WIKA operating instructions, model CPC2000
mmHg
■
Torr
■